dc.description.abstract |
Purpose: To report the first case of Neodymium Yttrium Aluminum Garnet (Nd:YAG) laser use to recanalize a
CyPass occluded with a clot in the early postoperative follow-up.
Observations: A 66 year-old woman with primary open angle glaucoma and cataract received combined cataract
surgery plus CyPass in the right eye. Surgery was uneventful, however, there was a small amount of hyphema
related to bleeding around the site of Cypass insertion. Intraocular pressure (IOP) was between 6 and 7 mmHg in
the first week on no glaucoma medications. At the second postoperative week, IOP jumped to 30 mmHg and a
clot was observed obstructing the CyPass lumen. We disrupted the clot with a single shot from a Nd:YAG laser.
Twenty minutes after the laser, the IOP was 8 mmHg. CyPass was recanalized without complications. IOP was
11 mmHg at 1 month postoperatively with timolol 0.5% twice a day.
Conclusions and importance: It is important to perform gonioscopy to evaluate for CyPass obstruction in cases of
IOP spike. Nd:YAG laser may be a useful approach to recanalize the CyPass in the early postoperative follow-up |
en_US |
dc.language.iso |
en |
en_US |
dc.publisher |
Elsevier Inc. |
en_US |
dc.relation.uri |
https://doi.org/10.1016/j.ajoc.2018.02.011 |
|
dc.rights |
Copyright © 2018 Elsevier |
en_US |
dc.title |
Use of Nd:YAG laser to recanalize occluded Cypass Micro-Stent in the early post-operative period |
en_US |
dc.type |
Article |
en_US |
dc.email.author |
No information provided |
|
dc.email.author |
No information provided |
|
dc.email.author |
No information provided |
|
dc.email.author |
No information provided |
|
dc.subject.keyword |
Glaucoma |
en_US |
dc.subject.keyword |
Lasers solid-state |
en_US |
dc.subject.keyword |
Cataract extraction |
en_US |