DSpace Repository

Use of Nd:YAG laser to recanalize occluded Cypass Micro-Stent in the early post-operative period

Show simple item record

dc.contributor.author Perez, Claudio I.
dc.contributor.author Sunee Chansangpetch
dc.contributor.author Hsia, Yen C.
dc.contributor.author Lin, Shan C.
dc.contributor.other Chulalongkorn University. Faculty of Medicine
dc.date.accessioned 2019-05-03T09:20:42Z
dc.date.available 2019-05-03T09:20:42Z
dc.date.issued 2018-06-01
dc.identifier.citation American Journal of Ophthalmology Case Reports, Vol. 10, (Jun 1, 2018) ; p. 114-116. en_US
dc.identifier.issn 2451-9936
dc.identifier.issn 10.1016/j.ajoc.2018.02.011
dc.identifier.uri http://cuir.car.chula.ac.th/handle/123456789/61686
dc.description.abstract Purpose: To report the first case of Neodymium Yttrium Aluminum Garnet (Nd:YAG) laser use to recanalize a CyPass occluded with a clot in the early postoperative follow-up. Observations: A 66 year-old woman with primary open angle glaucoma and cataract received combined cataract surgery plus CyPass in the right eye. Surgery was uneventful, however, there was a small amount of hyphema related to bleeding around the site of Cypass insertion. Intraocular pressure (IOP) was between 6 and 7 mmHg in the first week on no glaucoma medications. At the second postoperative week, IOP jumped to 30 mmHg and a clot was observed obstructing the CyPass lumen. We disrupted the clot with a single shot from a Nd:YAG laser. Twenty minutes after the laser, the IOP was 8 mmHg. CyPass was recanalized without complications. IOP was 11 mmHg at 1 month postoperatively with timolol 0.5% twice a day. Conclusions and importance: It is important to perform gonioscopy to evaluate for CyPass obstruction in cases of IOP spike. Nd:YAG laser may be a useful approach to recanalize the CyPass in the early postoperative follow-up en_US
dc.language.iso en en_US
dc.publisher Elsevier Inc. en_US
dc.relation.uri https://doi.org/10.1016/j.ajoc.2018.02.011
dc.rights Copyright © 2018 Elsevier en_US
dc.title Use of Nd:YAG laser to recanalize occluded Cypass Micro-Stent in the early post-operative period en_US
dc.type Article en_US
dc.email.author No information provided
dc.email.author No information provided
dc.email.author No information provided
dc.email.author No information provided
dc.subject.keyword Glaucoma en_US
dc.subject.keyword Lasers solid-state en_US
dc.subject.keyword Cataract extraction en_US


Files in this item

This item appears in the following Collection(s)

  • Foreign Journal Article [121]
    บทความวารสารภาษาต่างประเทศจากฐานข้อมูลออนไลน์ Free Open Access

Show simple item record